Consistency test of coincidence-summing calculation methods for extended sources
APPLIED RADIATION AND ISOTOPES, cilt.155, 2020 (SCI-Expanded, Scopus)
- Yayın Türü: Makale / Tam Makale
- Cilt numarası: 155
- Basım Tarihi: 2020
- Doi Numarası: 10.1016/j.apradiso.2019.108921
- Dergi Adı: APPLIED RADIATION AND ISOTOPES
- Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, Academic Search Premier, Aquatic Science & Fisheries Abstracts (ASFA), Chemical Abstracts Core, Chimica, Compendex, EMBASE, Food Science & Technology Abstracts, INSPEC, MEDLINE, Pollution Abstracts
- Anahtar Kelimeler: Gamma-ray spectrometry, HPGe detector, Coincidence-maiming corrections, Self-consistency test, GAMMA-RAY SPECTROMETRY, COMPUTER CODES, EFFICIENCY, SIMULATION, EQUIVALENCE, COMPUTATION, GESPECOR, TOOL
- Ankara Üniversitesi Adresli: Evet
Özet
An internal consistency test of the calculation of coincidence-summing correction factors F-C for volume sources is presented. The test is based on exact equations relating the values of F-C calculated for three ideal measurement configurations. The test is applied to a number of 33 sets of F-C values sent by 21 teams. Most sets passed the test, but not the results obtained using the quasi-point source approximation; in the latter case the test qualitatively indicated the magnitude of the bias of F-C.