COMBINED APPLICATION OF CONTINUOUS WAVELET TRANSFORM-ZERO CROSSING TECHNIQUE IN THE SIMULTANEOUS SPECTROPHOTOMETRIC DETERMINATION OF PERINDOPRIL AND INDAPAMID IN TABLETS


Pektas G., DİNÇ E., BALEANU D.

QUIMICA NOVA, cilt.32, sa.6, ss.1416-1421, 2009 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 32 Sayı: 6
  • Basım Tarihi: 2009
  • Dergi Adı: QUIMICA NOVA
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.1416-1421
  • Anahtar Kelimeler: continuous wavelet transform, zero crossing technique, perindopril, QUANTITATIVE RESOLUTION, ACETYLSALICYLIC-ACID, HUMAN PLASMA, MS METHOD, MIXTURE, SPECTRA, HYDROCHLOROTHIAZIDE
  • Ankara Üniversitesi Adresli: Evet

Özet

Signal processing methods based on the combined use of the continuous wavelet transform (CWT) and zero-crossing technique were applied to the simultaneous spectrophotometric determination of perindopril (PER) and indapamide (IND) in tablets. These signal processing methods do not require any priory separation step. Initially, various wavelet families were tested to identify the optimum signal processing giving the best recovery results. From this procedure, the Haar and Biorthogonal 1.5 continuous wavelet transform (HAAR-CWT and BIOR 1.5-CWT, respectively) were found suitable for the analysis of the related compounds. After transformation of the absorbance vectors by using HAAR-CWT and BIOR 1.5-CWT, the CWT-coefficients were drawn as a graph versus wavelength and then the HAAR-CWT and BIOR 1.5-CWT spectra were obtained. Calibration graphs for PER and IND were obtained by measuring the CWT amplitudes at 231.1 and 291.0 rim in the HAAR-CWT spectra and at 228.5 and 246.8 nm in BIOR 1.5-CWT spectra. respectively. In order to compare the performance of HAAR-CWT and BIOR 1.5-CWT approaches, derivative spectrophotometric (DS) method and HPLC as comparison methods, were applied to the PER-IND samples. In this DS method, first derivative absorbance values at 221.6 for PER and 282.7 nm for IND were used to obtain the calibration graphs. The validation of the CWT and DS signal processing methods was carried out by using the recovery study and standard addition technique. In the following step, these methods were Successfully applied to the commercial tablets containing PER and IND compounds and good accuracy and precision were reported for the experimental results obtained by all proposed signal processing methods.