COMPARATIVE EVALUATION OF MINIMUM DETECTABLE ACTIVITY CALCULATIONS FOR AN HPGE DETECTOR USING THE CURRIE METHOD AND ISO 11929 STANDARD
Türk Fizik Derneği 42. Uluslararası Fizik Kongresi, Muğla, Türkiye, 31 Ağustos - 04 Eylül 2026, ss.291, (Özet Bildiri)
- Yayın Türü: Bildiri / Özet Bildiri
- Basıldığı Şehir: Muğla
- Basıldığı Ülke: Türkiye
- Sayfa Sayıları: ss.291
- Ankara Üniversitesi Adresli: Evet
Özet
The Minimum Detectable Activity (MDA) is a fundamental performance parameter in
gammaray spectrometry, defining the lowest activity level that can be reliably distinguished
from background radiation. In this study, MDA values were calculated using both the
classical Currie method and the ISO 11929 standard for a high-purity germanium (HPGe)
detector and the results were comparatively evaluated. Measurements were performed using
an ORTEC HPGe detector with a relative efficiency of 78,5%, and gamma-ray spectra were
acquired and analyzed using GammaVision software. The calculations were based on
background counts, full-energy peak efficiencies, gamma emission probabilities, and
counting times. The Currie method. determines the critical level and detection limit using
statistical counting theory, whereas ISO 11929 incorporates the propagation of measurement
uncertainties, including counting statistics and calibration-related uncertainty components,
resulting in a more comprehensive estimation of the detection capability. MDA values were
calculated for selected gamma-ray energies, and the results obtained by the two approaches
were compared. The analysis showed that MDA values calculated according to ISO 11929
were generally higher than those obtained using the Currie method. The difference became
more pronounced in low-counting-statistics regions, reflecting the influence of uncertainty
contributions considered by the ISO standard. The results indicate that the Currie method
provides a simple and practical approach for routine measurements, while ISO 11929 offers
a more realistic and reliable assessment when measurement uncertainty must be rigorously
evaluated, particularly in environmental radioactivity measurements and accredited
laboratory applications. This comparative study provides guidance for selecting an
appropriate MDA calculation method in HPGe gamma-ray spectrometry.