COMPARATIVE EVALUATION OF MINIMUM DETECTABLE ACTIVITY CALCULATIONS FOR AN HPGE DETECTOR USING THE CURRIE METHOD AND ISO 11929 STANDARD


Demirel İ.

Türk Fizik Derneği 42. Uluslararası Fizik Kongresi, Muğla, Türkiye, 31 Ağustos - 04 Eylül 2026, ss.291, (Özet Bildiri)

  • Yayın Türü: Bildiri / Özet Bildiri
  • Basıldığı Şehir: Muğla
  • Basıldığı Ülke: Türkiye
  • Sayfa Sayıları: ss.291
  • Ankara Üniversitesi Adresli: Evet

Özet

The Minimum Detectable Activity (MDA) is a fundamental performance parameter in

gammaray spectrometry, defining the lowest activity level that can be reliably distinguished

from background radiation. In this study, MDA values were calculated using both the

classical Currie method and the ISO 11929 standard for a high-purity germanium (HPGe)

detector and the results were comparatively evaluated. Measurements were performed using

an ORTEC HPGe detector with a relative efficiency of 78,5%, and gamma-ray spectra were

acquired and analyzed using GammaVision software. The calculations were based on

background counts, full-energy peak efficiencies, gamma emission probabilities, and

counting times. The Currie method. determines the critical level and detection limit using

statistical counting theory, whereas ISO 11929 incorporates the propagation of measurement

uncertainties, including counting statistics and calibration-related uncertainty components,

resulting in a more comprehensive estimation of the detection capability. MDA values were

calculated for selected gamma-ray energies, and the results obtained by the two approaches

were compared. The analysis showed that MDA values calculated according to ISO 11929

were generally higher than those obtained using the Currie method. The difference became

more pronounced in low-counting-statistics regions, reflecting the influence of uncertainty

contributions considered by the ISO standard. The results indicate that the Currie method

provides a simple and practical approach for routine measurements, while ISO 11929 offers

a more realistic and reliable assessment when measurement uncertainty must be rigorously

evaluated, particularly in environmental radioactivity measurements and accredited

laboratory applications. This comparative study provides guidance for selecting an

appropriate MDA calculation method in HPGe gamma-ray spectrometry.