A temperature sensor implant for active implantable medical devices for in vivo subacute heating tests under MRI


Creative Commons License

Silemek B., Acikel V., OTO Ç., Alipour A., Aykut Z. G., ALGIN O., ...Daha Fazla

MAGNETIC RESONANCE IN MEDICINE, cilt.79, sa.5, ss.2824-2832, 2018 (SCI-Expanded) identifier identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 79 Sayı: 5
  • Basım Tarihi: 2018
  • Doi Numarası: 10.1002/mrm.26914
  • Dergi Adı: MAGNETIC RESONANCE IN MEDICINE
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.2824-2832
  • Anahtar Kelimeler: implant heating, active implantable medical devices, in vivo heating, heating reduction, parallel transmission, Internet of Things, DEEP BRAIN-STIMULATION, CARDIAC-PACEMAKER, INDUCED CURRENTS, 1.5 T, SAFETY, LEAD, SYSTEMS, FIELDS, WIRES
  • Ankara Üniversitesi Adresli: Evet

Özet

PurposeTo introduce a temperature sensor implant (TSI) that mimics an active implantable medical device (AIMD) for animal testing of MRI heating. Computer simulations and phantom experiments poorly represent potential temperature increases. Animal experiments could be a better model, but heating experiments conducted immediately after the surgery suffer from alterations of the thermoregulatory and tissue properties during acute testing conditions. Therefore, the aim of this study was to introduce a temperature sensor implant that mimics an AIMD and capable of measuring the electrode temperature after implantation of the device without any further intervention at any time after the surgery in an animal model.