A case report of an acute reversible increase in pacing impedance of a ventricular epicardial lead after catheter ablation of ventricular tachycardia


Baskovski E., CANDEMİR B., Akbulut M., ESENBOĞA K.

TURK KARDIYOLOJI DERNEGI ARSIVI-ARCHIVES OF THE TURKISH SOCIETY OF CARDIOLOGY, cilt.49, sa.2, ss.162-166, 2021 (ESCI) identifier identifier identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 49 Sayı: 2
  • Basım Tarihi: 2021
  • Doi Numarası: 10.5543/tkda.2021.79696
  • Dergi Adı: TURK KARDIYOLOJI DERNEGI ARSIVI-ARCHIVES OF THE TURKISH SOCIETY OF CARDIOLOGY
  • Derginin Tarandığı İndeksler: Emerging Sources Citation Index (ESCI), Scopus, Central & Eastern European Academic Source (CEEAS), EMBASE, MEDLINE, Directory of Open Access Journals, TR DİZİN (ULAKBİM)
  • Sayfa Sayıları: ss.162-166
  • Anahtar Kelimeler: Impedance, ventricular tachycardia, catheter ablation, artificial cardiac pacemaker, PACEMAKERS
  • Ankara Üniversitesi Adresli: Evet

Özet

Pacemakers are lifesaving devices that are being implanted with various indications, such as sinus node disease, atrioventricular block, and cardiac resynchronization therapy. Impedance measurement is one of the integral tests by which electrical resistance in pacing lead is tested. In this paper, we report an interesting observation of sudden impedance rise after ventricular tachycardia ablation with transmural lesions, which subsequently normalized without any need for intervention. The clinical implication of our findings is that careful observation might be indicated instead of lead replacement in case of a sudden surge in epicardial lead impedance after endocardial ablation owing that the impedance surge might be reversible.