Dependence of photon interaction depth on linear attenuation coefficient in high pure germanium detectors


YÜCEL H., Cetiner M., Demirel H.

APPLIED RADIATION AND ISOTOPES, cilt.47, sa.5-6, ss.535-541, 1996 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 47 Sayı: 5-6
  • Basım Tarihi: 1996
  • Doi Numarası: 10.1016/0969-8043(95)00325-8
  • Dergi Adı: APPLIED RADIATION AND ISOTOPES
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.535-541
  • Ankara Üniversitesi Adresli: Hayır

Özet

The present work is concerned with the construction of a semiempirical function d(e),(mu, E(gamma)) of the effective photon interaction depth depending upon both the linear attenuation coefficient of Ge and gamma-ray energy. The interaction depth results obtained from the measurements in two high purity germanium crystals, which are nearly equal in volume, have been fitted by using a function. The semiempirical function agrees with the measured interaction depth Values on the average to within 3-4%. The knowledge of effective interaction depth of gamma-rays in germanium detector is a useful parameter in order to calculate the absolute efficiency for any source-to-detector distance on the particular detector.