Torque-fitting and resonance frequency analyses of implants in conventional sockets versus controlled bone defects in vitro
International Journal of Oral and Maxillofacial Surgery, cilt.39, sa.2, ss.169-173, 2010 (SCI-Expanded, Scopus)
- Yayın Türü: Makale / Tam Makale
- Cilt numarası: 39 Sayı: 2
- Basım Tarihi: 2010
- Doi Numarası: 10.1016/j.ijom.2009.11.019
- Dergi Adı: International Journal of Oral and Maxillofacial Surgery
- Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
- Sayfa Sayıları: ss.169-173
- Anahtar Kelimeler: resonance frequency analysis, insertion torque, implant, iliac crest, bone defect, ENDOSSEOUS IMPLANTS, HUMAN CADAVER, STABILITY, EXTRACTION, PLACEMENT, INTERFACE, SITES, DOG
- Ankara Üniversitesi Adresli: Evet
Özet
The primary stability of implants should be high on insertion into fresh extraction sockets. Torque-fitting and resonance frequency analyses (RFA) are used to assess primary implant anchorage and stability. The torque-fitting and RFA of implants placed in conventional surgical sockets and sockets with controlled coronal bone defects was compared. The possible relation between torque-fitting and RFA was explored. Ø 3.3 mm × 12 mm implants were placed in 16 sockets finalized with Ø 2.8 mm surgical pilot drills in the right iliac crests of two fresh cadavers (control). In the test group, implants were placed into sockets prepared by Ø 2.8 mm drill followed by Ø 4.2 mm twist drills to a depth of 6 mm to create circumferential controlled coronal bone defects (50% bone loss). Primary implant stability was assessed using insertion torque values (ITV) followed by RFA. Mean ITV and RFA measurements for test groups (7.83 ± 0.91 N cm and 40.88 ± 3.57) were significantly lower than controls (14.80 ± 1 N cm and 66.31 ± 0.9) (P < 0.05). Reductions of ITV and RFA measurements in relation to bone defect were 47% and 38%. The existence of controlled bone defects eliminating contact coronally leads to decrease in torque-fitting and primary stability of implants. No relationship was observed between torque-fitting and RFA. © 2009 International Association of Oral and Maxillofacial Surgeons.