MONTE CARLO-BASED FORWARD MODELING AND RESPONSE SURFACE GENERATION FOR URANIUM ENRICHMENT ANALYSIS USING HPGE GAMMA SPECTROMETRY


Demirel İ.

Türk Fizik Derneği 42. Uluslararası Fizik Kongresi, Muğla, Türkiye, 31 Ağustos - 04 Eylül 2026, ss.292, (Özet Bildiri)

  • Yayın Türü: Bildiri / Özet Bildiri
  • Basıldığı Şehir: Muğla
  • Basıldığı Ülke: Türkiye
  • Sayfa Sayıları: ss.292
  • Ankara Üniversitesi Adresli: Evet

Özet

A Monte Carlo based forward modeling framework is developed for uranium enrichment

analysis using high-resolution gamma spectrometry. A high-purity germanium (HPGe)

detector is modeled with detailed geometry, and its energy response is implemented via a

Gaussian energy broadening (GEB) function derived from experimentally measured full

width at half maximum (FWHM) data. The proposed approach introduces a hybrid

methodology in which full-energy peak areas are obtained from Monte Carlo simulations,

while full-energy peak efficiencies are transferred from experimentally calibrated reference

measurements with differing matrix composition. Gamma emission probabilities are taken

from evaluated nuclear data libraries. The analysis focuses on characteristic gamma emissions

of 235U at 185.7 keV and the 1001 keV transition of 234mPr associated with the decay

representation of 238U.Unlike conventional assay approaches, the framework is formulated

as a forward response model rather than a direct enrichment measurement method. It

decouples detector response modeling from experimental counting statistics to a controlled

extent, enabling systematic investigation of enrichment-dependent spectral behavior under

idealized simulation conditions. Monte Carlo simulations are performed for multiple

enrichment scenarios ranging from natural uranium to 10% enrichment, generating a detector

response surface based on efficiency-corrected peak areas. The methodology is validated by

reproducing the expected activity ratio for natural uranium (~21.5), demonstrating internal

consistency of the detector model and efficiency transfer scheme. Uncertainty contributions

arising from efficiency transfer, statistical fluctuations in Monte Carlo sampling, and weak

gamma line statistics are identified as the dominant sources of systematic deviation. The

results show that the proposed framework provides a strong forward modeling and response

surface approach for uranium enrichment studies. It improves the interpretability of detector

response effects and establishes a validated computational basis for future inverse problem

formulations in nuclear spectrometry applications.