YÜCEL H., Cubukcu S., Uyar E., Engin Y.
PHYSICS IN MEDICINE AND BIOLOGY, cilt.59, sa.22, ss.6749-6758, 2014 (SCI-Expanded)
-
Yayın Türü:
Makale / Tam Makale
-
Cilt numarası:
59
Sayı:
22
-
Basım Tarihi:
2014
-
Doi Numarası:
10.1088/0031-9155/59/22/6749
-
Dergi Adı:
PHYSICS IN MEDICINE AND BIOLOGY
-
Derginin Tarandığı İndeksler:
Science Citation Index Expanded (SCI-EXPANDED), Scopus
-
Sayfa Sayıları:
ss.6749-6758
-
Anahtar Kelimeler:
plastic scintillator, BC-408, energy response, medical dosimeter, beam quality, x-ray tube, DIAGNOSTIC-RADIOLOGY, DOSIMETRY
-
Ankara Üniversitesi Adresli:
Evet
Özet
The energy dependence of the response of BC-408 plastic scintillator (PS), an approximately water-equivalent material, has been investigated by employing standardized x-ray beams. IEC RQA and ISO N series x-ray beam qualities, in the range of 40-100 kV(p), were calibrated using a PTW-type ionization chamber. The energy response of a thick BC-408 PS detector was measured using the multichannel pulse height analysis method.