Silver(I)-selective PVC membrane potentiometric sensor based on a recently synthesized calix[4]arene
Electroanalysis, cilt.18, sa.10, ss.1019-1027, 2006 (SCI-Expanded, Scopus)
- Yayın Türü: Makale / Tam Makale
- Cilt numarası: 18 Sayı: 10
- Basım Tarihi: 2006
- Doi Numarası: 10.1002/elan.200503494
- Dergi Adı: Electroanalysis
- Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
- Sayfa Sayıları: ss.1019-1027
- Anahtar Kelimeler: potentiometry, ion-selective electrode, calix[4]arene, silver(I) ion, ION-SELECTIVE ELECTRODES, COATED-WIRE ELECTRODE, POLYMERIC MEMBRANES, NEUTRAL CARRIER, BULK OPTODES, SILVER, IONOPHORES, DERIVATIVES, SODIUM, ETHER
- Ankara Üniversitesi Adresli: Evet
Özet
A new PVC membrane potentiometric sensor for Ag(I) ion based on a recently synthesized calix[4]arene compound of 5,11,17,23-tetra-tert-butyl-25,27- dihydroxy-calix[4]arene-thiacrown-4 is developed. The electrode exhibits a Nernstian response for Ag(I) ions over a wide concentration range (1.0 × 10-2 - 1.0 × 10-6 M) with a slope of 53.8 ± 1.6 mV per decade. It has a relatively fast response time (5-10 s) and can be used for at least 2 months without any considerable divergence in potentials. The proposed electrode shows high selectivity towards Ag+ ions over Pb2+, Cd2+, Co2+, Zn2+, Cu 2+, Ni2+, Sr2+, Mg2+, Ca 2+, Li+, K+, Na+, NH 4+ ions and can be used in a pH range of 2-6. Only interference of Hg2+ is found. It is successfully used as an indicator electrode in potentiometric titration of a mixture of chloride, bromide and iodide ions. © 2006 Wiley-VCH Verlag GmbH & Co. KGaA.