EDTA modified glassy carbon electrode: Preparation and characterization


Üstündaǧ Z., Solak A. O.

Electrochimica Acta, vol.54, no.26, pp.6426-6432, 2009 (SCI-Expanded) identifier

  • Publication Type: Article / Article
  • Volume: 54 Issue: 26
  • Publication Date: 2009
  • Doi Number: 10.1016/j.electacta.2009.06.015
  • Journal Name: Electrochimica Acta
  • Journal Indexes: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Page Numbers: pp.6426-6432
  • Keywords: EDTA nanofilm, Electrochemical impedance spectroscopy, Ellipsometry, X-ray photoelectron spectroscopy
  • Ankara University Affiliated: Yes

Abstract

EDTA-phenoxyamide modified glassy carbon electrode (EDTA-GC) was prepared at a glassy carbon electrode by surface synthesis. In the first step, nitrophenyl was grafted to the glassy carbon (GC) surface via the electrochemical reduction of its tetraflouroborate diazonium salt. In the second step, nitrophenyl-modified electrode (NP-GC) was subjected to the cathodic potential scan to reduce the nitro to amine group. p-Aminophenyl modified glassy carbon electrode (AP-GC) was dipped into a EDTA solution containing 1-ethyl-3(3-(dimethlyamino)propyl)-carbodiimide (EDC) as an activating agent. Thus formed ((2-anilino-2-oxoethyl){2-[bis(carboxymethyl)amino]-ethyl}amino)acetic acid modified GC electrode was denoted as EDTA-GC and characterized by cyclic voltammetry (CV), electrochemical impedance spectroscopy (EIS), ellipsometry and X-ray photoelectron spectroscopy (XPS). Complexation of the EDTA-GC surface with Pb2+ ions was investigated if this electrode could be used as a metal sensor. © 2009 Elsevier Ltd. All rights reserved.