Salicylate Ion-Selective Electrode Based on a Calix[4]arene as Ionophore


Ayanoglu M. N., KORMALI ERTÜRÜN H. E., DEMİREL ÖZEL A., ŞAHİN Ö., YILMAZ M., Kilic E.

ELECTROANALYSIS, cilt.27, sa.7, ss.1676-1684, 2015 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 27 Sayı: 7
  • Basım Tarihi: 2015
  • Doi Numarası: 10.1002/elan.201400737
  • Dergi Adı: ELECTROANALYSIS
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.1676-1684
  • Anahtar Kelimeler: Salicylate-selective electrode, Calix[4]arene, Potentiometry, Salicylate determination, POTENTIOMETRIC MEMBRANE-ELECTRODE, TANDEM MASS-SPECTROMETRY, NEUTRAL CARRIER, LIQUID-MEMBRANE, SCHIFF-BASE, COPPER(II) COMPLEX, ACID, CALIXARENE, SENSOR, DERIVATIVES
  • Ankara Üniversitesi Adresli: Evet

Özet

A salicylate-selective electrode based on calix[4]arene derivative was developed and its response characteristics were investigated. The optimum membrane composition was 1% ionophore, 30% PVC, 69% DOS. The electrode exhibited a Nernstian slope of 58.8 +/- 0.5mV/pSal in the range of 1.0x10(-5)-1.0x10(-1)M with a detection limit of 4.3x10(-6)M at pH4.0, 20 +/- 1 degrees C. The potentiometric response of the electrode in the presence of different anions was investigated by the separate solution method. The lifetime was found at least 4months, and its response time was 5-10s. It was successfully used for the potentiometric determination of salicylate in pharmaceutical preparations.