Mechanical failure with malleable penile prosthesis


Akand M., Ozayar A., YAMAN M. Ö., Demirel C.

UROLOGY, vol.70, no.5, 2007 (SCI-Expanded) identifier identifier identifier

  • Publication Type: Article / Article
  • Volume: 70 Issue: 5
  • Publication Date: 2007
  • Doi Number: 10.1016/j.urology.2007.08.005
  • Journal Name: UROLOGY
  • Journal Indexes: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Ankara University Affiliated: No

Abstract

Mechanical failure with a malleable penile prosthesis is very rare, and no mechanical failure has been reported with the AMS 650 malleable penile Prosthesis. A 67-year-old man, using the AMS 650 malleable penile prosthesis without a problem since 1999, presented with right penile deformity and difficulty in sexual intercourse for the preceding 2 months. Physical examination showed dislocation of the right rod of the penile prosthesis, and surgical exploration revealed a fracture in the rod. The malfunctioning right penile prosthesis was removed, and a new AMS 650 malleable penile prosthesis was implanted. The patient did not have any problem at I-year follow-up.