Lensless Imaging for Point-of-Care Testing


Creative Commons License

Moon S., Keles H. O., Kim Y., Kuritzkes D., Demirci U.

Annual International Conference of the IEEE-Engineering-in-Medicine-and-Biology-Society, Minnesota, Amerika Birleşik Devletleri, 3 - 06 Eylül 2009, ss.6376-6377 identifier identifier

  • Yayın Türü: Bildiri / Tam Metin Bildiri
  • Doi Numarası: 10.1109/iembs.2009.5333765
  • Basıldığı Şehir: Minnesota
  • Basıldığı Ülke: Amerika Birleşik Devletleri
  • Sayfa Sayıları: ss.6376-6377
  • Ankara Üniversitesi Adresli: Hayır

Özet

We show a platform that merges a microfluidic chip with lensless imaging for CD4(+) T-lymphocyte counting at resource-limited settings. To capture CD4(+) T lymphocytes, anti-CD4 antibody was immobilized on a microfluidic chip. The captured cells were detected by a charge coupled device (CCD) sensor using lensless shadow imaging techniques. Gray scale shadow images of captured cells on the chip (24 mm x 4 mm x 50 mu m) were enumerated in three seconds using an automatic cell counting software. The device achieved 70.2 +/- 6.5% capture efficiency, 88.8 +/- 5.4% capture specificity for CD4(+) T-lymphocytes, 96 +/- 1.6% CCD efficiency, and 83.5 +/- 2.4% overall platform performance (n = 9 devices). This integrated platform has potential for point-of-care testing (POCT) to rapidly capture, image and count specific cell types from unprocessed whole blood.