Fully Validated Simultaneous Determination of Bisoprolol Fumarate and Hydrochlorothiazide in Their Dosage Forms Using Different Voltammetric, Chromatographic, and Spectrophotometric Analytical Methods


Bozal Palabıyık B., Gümüştaş M., Doğan Topal B., Uslu B., Özkan S. A.

JOURNAL OF AOAC INTERNATIONAL, vol.96, no.1, pp.42-51, 2013 (SCI-Expanded) identifier

  • Publication Type: Article / Article
  • Volume: 96 Issue: 1
  • Publication Date: 2013
  • Doi Number: 10.5740/jaoacint.11-364
  • Journal Name: JOURNAL OF AOAC INTERNATIONAL
  • Journal Indexes: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Page Numbers: pp.42-51
  • Ankara University Affiliated: Yes

Abstract

Voltammetric, chromatographic, and spectrophotometric methods were developed for the simultaneous determination of bisoprolol fumarate (BIS) and hydrochlorothiazide (HCZ). Differential pulse and square wave voltammetry techniques were used to analyze BIS and HCZ simultaneously by measuring at about 1400 and 1100 mV, respectively. RP-HPLC was the second method for simultaneous analysis of the compounds. The mixture of BIS, HCZ, and moxifloxacin as an internal standard was separated on an RP Zorbax Eclipse XDB-C18 column (150 x 4.6 mm, id, 5 mu m particle size) using acetonitrile-15 mM phosphate (25+75, v/v) mobile phase at a 1.0 mL/min flow rate. The third method was based on first derivative of the ratio-spectra method obtained from the measurements of the amplitudes at 246 and 257 nm for BIS and HCZ, respectively. All the proposed methods were effectively applied for the simultaneous determination of BIS and HCZ in tablet dosage forms without any time-consuming extraction, sample preparation, or derivatization procedures.