BIOMED RESEARCH INTERNATIONAL, cilt.2018, 2018 (SCI-Expanded)
Objectives. To examine the effect of fluoroscopy devices with different sizes of image intensifier and C-arm maneuverability on operating time, fluoroscopy time, radiation dose and reduction, and fixation quality at intertrochanteric femoral fractures. Design. Single-center, randomized, prospective study. Setting. Academic Level I trauma hospital. Patients and Intervention. 34 patients treated with cephalomedullary nailing for a stable, intertrochanteric proximal femur fracture (OTA A1). Main Outcome Measurement. The total working time of the fluoroscopy device, the dose-area product (DAP), operating time, reduction quality (cortical continuity, symmetrical collodiaphyseal angle, and shortness), and fixation quality (Bosworth quadrants, the tip-apex distance, TAD). Results. There were no cases of poor reduction; also the placement of the blade was optimal for 14 patients and suboptimal in 3 patients in each group. Superior-posterior placement of the blade or TAD > 25 mm was not seen in any patient. Total operating time was significantly shorter when using device A compared to the use of device B (20.1 +/- 3.4 mins versus 25.3 +/- 5.4 mins, p < 0.001). Total radiation time was significantly shorter with device A compared to the use of device B (58.1 +/- 19.4 secs versus 98.9 +/- 55.4 secs, p = 0.008). The measured radiation dose was lower with the use of device A compared to device B (3.5 +/- 1.2 Gy.cm(2) versus 7.3 +/- 4.5 Gy.cm(2), p = 0.002). Conclusion. Physical properties of fluoroscopy devices used during the fixation of intertrochanteric fractures could yield significant differences in operating times and the radiation dose while having comparable clinical results.