EFFICIENCY CALIBRATION OF A WELL-TYPE Ge DETECTOR FOR VOLUMINOUS SAMPLES IN CYLINDRICAL GEOMETRY


Solmaz A. N., Bor D., Yuecel H.

NATO Advanced Training Course in New Techniques for the Detection of Nuclear and Radioactive Agents, Muğla, Turkey, 26 - 30 May 2008, pp.193-194, (Full Text) identifier identifier

  • Publication Type: Conference Paper / Full Text
  • Doi Number: 10.1007/978-1-4020-9600-6_17
  • City: Muğla
  • Country: Turkey
  • Page Numbers: pp.193-194
  • Keywords: Radiation detection, well-type Germanium detectors, SELF-ABSORPTION CORRECTIONS, GAMMA-RAY SPECTROMETRY
  • Ankara University Affiliated: Yes

Abstract

To study environmental samples, a 44.8% relative efficient well-type Ge detector with all active volume of 218 cm(3) was calibrated ill three different voluminous sample geometries. The models were tested by absolute full energy peak efficiency calibration with use of certified multinuclide standards. The efficiency results corrected for this experiment obtained self-absorption factors from the simple photon transmission experiments by using point-gamma sources for small cylindrical geometry and theoretically calculated self absorption factors based on the database related to mass attenuation coefficients for vial geometry. Tufa, soil, and marble were also measured at these counting geometries. The activity results obtained with these models were compared with each other.